OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

Our eternal pursuits are the attitude of "regard the market, regard the custom, regard the science" as well as the theory of "quality the basic, believe in the 1st and management the advanced" for Water Monitoring Tds Sensor , 4-20ma Pressure Sensor Transmitter , Orp Ph Meter , You would not have any communication problem with us. We sincerely welcome customers all over the world to contact us for business cooperation.
OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS Detail:

Pls unpack and check that the sensor has been supplied undamaged and that it is the correct option as ordered. If you have any problems please contact your supplier.

Application
Widely used for industrial water, tap water, cooling water, pure water etc. Conductivity measurement.

Main technique specification

Name

Function

Cell Constant

0.05cm-1  ( v )

0.1cm-1  (  )

1.0cm-1  (  )

10.0cm-1 (  )

Electrode structure

Bipolar

Electrode material

ABS (   )

316L Stainless steel(  v )

Temperature Sensor

NTC 10K ( v )

Pt 1000  (  )

 Pt 100  (  ) 

Thread structure

½” NPT Thread

Working pressure

0~0.5MPa

Operating temperature

0~50℃

Cable length

Standard:5m or others(5-30m)

Dimensions drawing
Conductivity-meter5

Conductivity(TDS)/ Resistivity electrode dimensions 

Installation and maintenance
Installation: To ensure the real measurement result, data distortion caused by air bubble or dead water in conductance cell should be avoided. The installation should be performed strictly according to the following drawing:

Conductivity-meter6

Notes
1. The electrode should be installed at a lower location in the pipe where the flow speed is steady and air bubbles are seldom generated.
2. No matter the conductance cell is horizontally or vertically installed, it should be deeply inserted into the moving water.
3. The conductivity/Resistivity signal is weak electronic signal and its collecting cable should be separately installed.
When threading cable joint or connecting terminal board is used, to avoid wetting interference or breakdown of measurement unit circuit, they should not be connected to the same group of cable joint or terminal board with the power or control line.
4. When the measurement cable needs to be lengthened, it’s recommended to use the cable provided by the original manufacturer, and the joint should be subject to reliable damp-proofing insulation disposal. When a longer distance is involved, the length of the cable (<30m) should be agreed on before delivery, and if the length is over 30m, a transmitter should be used.

Electrode Maintenance
1. The electrode cell should not be soaked in strong acid or alkali liquid, and the platinum black coating should not be wiped or it will lead to electrode surface damage and the constant and response capability will be affected. The right way should be: when the electrode is dirty, soak it in 10% dilute hydrochloric acid for a short while, then rinse it with pure water to keep the surface clean.
2. The measurement cable is special cable and should not be changed at will or it will cause significant error.

Joint wire
white wire to Cell -INPUT
Yellow wire to Cell -OUPUT
Black wire-TEMP
Red wire-TEMP

JiShen Water treatment Co., Ltd.
Add: No.18, Xingong Road, High-Technology Area, Shijiazhuang,China
Tel: 0086-(0)311-8994 7497 Fax:(0)311-8886 2036
E-mail:info@watequipment.com
Website: www.watequipment.com


Product detail pictures:

OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS detail pictures

OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS detail pictures

OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS detail pictures


Related Product Guide:

We normally think and practice corresponding towards the change of circumstance, and grow up. We aim at the achievement of a richer mind and body plus the living for OEM/ODM China Tds Sensor 4-20ma - Online Conductivity TDS Sensor CR-102P – JIRS , The product will supply to all over the world, such as: Panama, Norwegian, Malta, Qualified R&D engineer will be there for your consultation service and we will try our best to meet your requirements. So please feel free to contact us for inquiries. You'll be able to send us emails or call us for small business. Also you are able to come to our business by yourself to get further knowing of us. And we will surely give you the best quotation and after-sale service. We're ready to build stable and friendly relations with our merchants. To achieve mutual success, we will make our best efforts to build a solid co-operation and transparent communication work with our companions. Above all, we are here to welcome your inquiries for any of our goods and service.
  • Goods just received, we are very satisfied, a very good supplier, hope to make persistent efforts to do better.
    5 Stars By Grace from Greek - 2018.02.04 14:13
    In our cooperated wholesalers, this company has the best quality and reasonable price, they are our first choice.
    5 Stars By Edwina from Eindhoven - 2018.12.25 12:43
    Write your message here and send it to us